GF92 Installation, Operation and
Troubleshooting Guide
Document 06EN003321 Rev. -24 Hour Factory Service Hot Line: 1 (800) 854-1993
A. To get the best results from the instrument, the sensor should be located 20 pipe diameters downstream from any
flow disturbance (valve, pipe elbow, etc.) and 10 pipe diameters upstream from any disturbance.
The instrument tags show the model number, tag number (if noted on the customers order), serial number along
with other important safety information. Compare this information with the appropriate pipe installation drawings
and calibration sheets to verify the instrument is the correct configuration.
Tag Location - Between Conduit Ports
On Local Enclosure
Tag Location - Top Side of
Remote Enclosure
Flow Element Serial Number
Also Showing Flow Arrow
(Located near the FE enclosure.
It is also on the enclosure tag.)
B. Verify the serial numbers on the enclosure(s), flow element and electronics match. The instrument may not work if the
serial numbers do not match. (The instrument has a remote transmitter enclosure (FT) and a local flow element
enclosure (FE).)
(Top circuit board, lower left corner. It is also on
the transmitters (FT) enclosure tag.
Electronics Serial Number
If the instrument is a remote configuration, the serial number on the enclosure tags must match. The recommended tag
number on the local enclosure will have an FT in the tag number. The recommended tag number on the remote enclosure
will have an FE in the tag number. (Tags are specified by the customer, “FE/FT” is a recommended naming convention .)
C. Recommended installation/troubleshooting tools are an open-ended wrench to fit the NPT connection, an open-
ended wrench to fit the flanged fitting nuts and bolts, a small flat blade screw driver for manipulating potentiometers,
both a medium flat blade screwdriver and a medium phillips head screwdriver for tightening connections, 3 mm allen
wrench for CENELEC approved instruments, a measuring tape for proper flow element placement, and a DVM for
Ohm/Voltage measurements.
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - Troubleshooting Guide

GF92 Installation, Operation andTroubleshooting GuideDocument 06EN003321 Rev. -24 Hour Factory Service Hot Line: 1 (800) 854-1993A. To get the best r

Page 2

Model GF92 Flowmeter 2 Doc. No. 06EN003321 Rev.-Step 2. Flow Element InstallationIn-Line GF92 Flow Element and Enclosure Local NEMA 4 &7 Enclos

Page 3 - Step 3. Wiring Preparation

Doc. No. 06EN003321 Rev. -3 Model GF92 FlowmeterINSTALLATION, OPERATION AND TROUBLESHOOTING GUIDE FLUID COMPONENTS INTLStep 3. Wiring PreparationBefo

Page 4

Model GF92 Flowmeter 4 Doc. No. 06EN003321 Rev.-Analog Output 2 is connected in a similar manor as Analog Output 1. ( For Voltage Output: 0 - 5, 0

Page 5 - 5 Model GF92 Flowmeter

Doc. No. 06EN003321 Rev. -5 Model GF92 FlowmeterINSTALLATION, OPERATION AND TROUBLESHOOTING GUIDE FLUID COMPONENTS INTL..Wiring the Flow Element:Conne

Page 6 - Step 5. Operation

Model GF92 Flowmeter 6 Doc. No. 06EN003321 Rev.-The instrument has been configured and calibrated to custom specifications. In-depth programming o

Page 7 - Normal Operation

Doc. No. 06EN003321 Rev. -7 Model GF92 FlowmeterINSTALLATION, OPERATION AND TROUBLESHOOTING GUIDE FLUID COMPONENTS INTLQUICK OPERATIONMENU0.0 MAIN MEN

Page 8 - Notice of Proprietary Rights

Model GF92 Flowmeter 8 Doc. No. 06EN003321 Rev.-If there are still problems with the instrument, see the Troubleshooting Section in the GF Series M

Comments to this Manuals

No comments